Class B Type Automatic Horizontal Pulsating Vacuum Sterilizer For Hospotal

Sada Medical Class B Type Automatic Horizontal Pulsating Vacuum Sterilizer For Hospotal

Sada Medical Class B Type Automatic Horizontal Pulsating Vacuum Sterilizer For Hospital

Model: MSG-B
Sterilizing Volume: 60L, 80L, 120L, 140L, 180L

Application:

MSG series The full-automatic horizontal sterilizer is applicable to the sterlization of the articles from the depart-ments of the stomatology,the ophthalmology, the operating room, the laboratory and the other places.The full-automatic horizontal pulsating vacuum sterilizer can be used for the sterilization of theMSG.B serieshigh-temperature resistant and high-pressure resistant articles, such as the implantable instruments that are pack-aged, non-packaged, solid, Class A hollow,Class B hollow,and the instruments with the pipes inside.MSG.N series The full-automatic horizontal downdraft sterilizer(non-pulsating) can sterilize the exposed instru-ments,the packages of instruments, the packages of dressings,the rubber products,the liquids, the test instrumentsand the other articles resistant to the high temperature and the high pressure.

Shipping Port:Shanghai, Qingdao, Ningbo, Shenzhen, Guangzhou, etc.
Lead Time:20 – 25 working days.

Features of Horizontal Pressure Steam Sterilizer

The control system adopts the PLC,and the man-machine interface adopts the color touch screen of 65 thousand.The whole system has powerful functions and high reliability.

The whole process,including water injection, pulsation (unique to the pulsating vacuum sterilizers), heatima, steril-ization and drying, runs automatically without manual supervision.

The touch screen can display the information of the status, the temperature, the pressure,the time anid thie “aultalarm in real time.

Equipped with a micro printer, can record all of the information of the whole sterilization process.

It has the function of periodic counting. After the sterilization cycle is completed,there will be the text display on the touch screen and the buzzing prompts.

MSG.B series The full automatic horizontal pulsating vacuum sterilizer has a built-in high-quality water ring pump,pulsating vacuum for 1-6 times, and the vacuum degree up to -0.09MPa (-0.9Bar),so to ensure the efficient removalof the air in the dressing and the disinfection bags and the gaps between the central control devices, in order toachieve the best vaporous saturation.

MSG.B seriesThe full-automatic horizontal vacuum sterilizer adopts the internal circulation structure.and thewater ring pump does not need the external pipeline, which simplifies the installation requirements.MSG.B series The ful-automatic horizontal vacuum sterilizer has a unique program of pulsating vacuum drying.The residual humidity after drying does not exceed 0.2%.

Safety Protection

The protective device for the heater:The system cuts off the power of the heater automatically and displays the alarm information,when the heater exceeds the set temperature

The alarm device for preventing dry-burning: The anti-dry burning device cuts off the power supply of the heater automatically and displays the alarm information,when the water level is lower than the heating pipe;

The door safety interlocking device (volume 80L):it uses an electronic door lock. The sterilizer can start the working procedure,only when the door is closed in place.The door lock is automatically closed and door can not be opened,when there is pressure in the cavity or the power supply is not turned on;

The door safety interlocking device (The volume> 120L):Only when the door is locked in place can the sterilizer start to work.The door safety interlocking device will be locked and the door can not be opened when there is pressure in the chamber;

The protective device for overtemperature and overpressure:The system alarms,the touch screen displays the alarm information and the program is interrupted automatically,when the temperature and the pressure inside the sterilization chamber exceed the safety range.Then the pressure is released and cooled to protect the articles to be sterilized from being damaged;

The safety valves for automatic pressure relief,when overpressure occurs:The safety valve is opened and the pressure is released,when the pressure in the sterilizing chamber exceeds the set pressure;

The protective measures for the electric power:The design has the protective function for short circuit and leakage to ensure the safety of the equipment and the person.

Specifications

Brand: Sada Medical

Model: MSG-B
Sterilizing Volume: 60L, 80L, 120L, 140L, 180L

The main technical parameters:

Design Pressure: 0.26mpa,  Design temperature: 150ºC

Maximum working pressure: 0.23mpa,  The accuracy of temperature display: 0.1ºC

Selection of sterilization temperature: 105-138ºC

Uniformity of temperature: ≤±0.5ºC

Material of the cavity: SU304 stainless steel

Requirements for water sources: Distilled water or purified water

Model Capacity Inner size(ΦxL)mm Overall dimension(L x W x H)mm Net Weight(kg) Voltage Frequency power
MSG-B60 60L 340X680 780x480x1450 175 220v 50hz 7.6kw
MSG-B80 80L 370X755 870x540x1450 185 380v 50hz 7.6kw
MSG-B120 120L 500×665 880x650x1550 252 380v 50hz 10.2kw
MSG-B140 140L 500×765 880x650x1550 275 380v 50hz 10.2kw
MSG-B180 180L 500×910 1180x655x1550 360 380v 50hz 10.5kw